Categories:

    Name (required)

    Email (required)

    Phone (required)

    Country (required)

    Product

    Message

    Genlab is proud to offer their customers a unique range of sustainable laboratory products and services. E3 is our market leading brand for scientifically developed, cutting edge, sustainable, eco-friendly products. Genlab’s new E3 range of glassware drying cabinets are energy efficient, safer and economic to run. Working with the University of Cambridge, we developed units that represent a novel, sustainable solution to glassware drying. The feature our latest touch screen control system, which offer intuitive control and excellent accuracies with bespoke firmware for energy saving control, including adjustable automatic on/off times.

    Design

    The exterior is constructed from sheet steel finished in an easy clean powder coated paint. The interior chamber is made from 304 stainless steel and all units have high density insulation. The 100 and 200 litre models have sliding glass doors and the 425 and 885 litre models have double glazed hinged doors.

    Heating

    Heated by Incoloy sheathed elements which are positioned in the lower chamber and covered with 304 stainless steel guard.

    Controls

    The control system comprises of a bespoke touch screen interface that offers both accurate temperature control and integral overheat system by using 2 individual Pt100 sensors. The cabinets automatically turn on and off (up to two times per day) with boost and extended functions available outside of the set times. Oven trending is displayed for up to 48 hours.

    We have created a tutorial video that demonstrates how to use the touch screen control panel of the E3 units. This can be found by clicking here to access the video website page & it is listed as “Chapter 8 Genlab E3 Range Operation”

    Additional information

    Litre Size

    100, 885

    Features

    • Temperature range 30.0 to 80.0ºC
    • Integral 7 day timer with touch screen digital control
    • Integral energy meter
    • Fully insulated with adjustable vent cover
    • Stainless steel chamber
    • Manual or automatic overheat reset
    • On screen historical trending (48 hours)
    • Lockable castors on the 425 and 885 litre models
    • Low energy consumption – More than 50% lower to traditional cabinets
    • Low heat output – reducing air conditioning cost
    • High accuracy Pt100B duplex sensors <0.8ºC
    • Excellent stabilty <+/- 0.6ºC
    • New – Set point lock feature

    Options

    • Audible warnings
    • Access ports (25, 50, 75 or 100mm)
    • Traceable calibration to national standards
    • Bespoke stands and stacking kits
    • Wall mounting brackets (100 and 200 versions only)
    • Extractor unit
    • Extended warranty
    • Bespoke solutions available upon request

    Specifications

    Fan Assisted
    Genlab ref.
    Number
    Capacity
    (litres)
    Internal dims.
    (H x W x D cms)
    External dims.
    (H x W x D cms)
    No. of
    Shelves
    Wattage
    /Voltage
    Energy Consumption
    kWh/day*
    Weight
    (kg)
    E3DWC100F/TDIG
    100
    40 x 67 x 37
    66 x 72 x 42
    3..4
    500…230
    8.35
    52
    E3DWC200F/TDIG
    200
    49 x 93 x 45
    77 x 100 x 50
    3..4
    750…230
    12.97
    77
    E3DWC425F/TDIG
    425
    135 x 53 x 59
    175 x 60 x 65
    3..12
    1750…230
    18.32
    142
    E3DWC885F/TDIG
    885
    135 x 111 x 59
    175 x 118 x 65
    3..12
    2500…230
    27.18
    210